Model Number: SSI 567 FTISO
Standard Number: ISO 15025
The ISO test serves as a preliminary indication of Textile / plastics material acceptability for use as part of a device or appliance with respect to its flammability. The test is conducted in a Five sided chamber without the roof top.
|
The ISO test serves as a preliminary indication of Textile / plastics material acceptability for use as part of a device or appliance with respect to its flammability.
This test uses specimen is pinned to the sample holder in vertical position at 300mm from the nearest wall. A micro burner is used to apply flame at 30o, 45o or 90o for specific time.
The following parameters are then observed
The instrument conforms to ISO: 15025
The Instrument consist of the following
A five sided Chamber for sample burning with front glass door
A adjustable Burner with precise gas control system including gas regulator
Burner Moving system to adjust the burner position
Sample Holding System
Digital Timer for flame application (can be preset for 1 to 99 sec) to exhaust the burner the end of desired time
Digital timer for flame observation after flame is removed.
The timer starts automatically once the flame application time is up.
Digital timer for Glow observation the timer starts automatically once the after flame timer is stopped .