•  +91 9811209579
  • info@ssinstruments.com
  •  S.S. Instruments Pvt. Ltd.

       Material Testing Instruments
ISO 9001:2008 Certifited   
8 cell photo s

Ageing Oven 8 Cell


Model Number: SSI 502 AGISFA8C

Standard Number: IS: 5831IS: 10810 (Pt.10&11)IEC 811-1-2IEC 8011-3-2 (section 8)UL 2556

Thermoplastic and Poly vinyl Chloride substances when exposed to heat undergo many types of physical and chemical changes. The extent and type of change that takes place depends upon the severity of exposure of the sample to heat (i.e., temp. range), duration and rate of air flow

 

Salient Features
Model :  SSI 502 AGISFA8C (Ageing Oven as per IS & IEC , Flexible Air setting,  8 Cell Model)
Chamber Size :  100mm (Dia) x 300 mm (Height)
No of Chambers :  8 chambers, Independently controlled
Instrument Construction :  Test Chamber is made of thick Aluminum Pipe, MS / GI outer Chamber
Temperature Range :  Ambient to 200oC, LC 0.1oC
Temperature Control  Using PID Digital Temperature Controller, +/- 2oC
Air Flow  0-1 LPM / 0-10 LPM (Interchangeable) , Contolled using Needle Valves
Air Flow Display :  Using Digital Flow Sensors
Instrument Control Individual Control for Temperature, combined control for Air
Delivery :  4 to 6 Weeks
Packing :  One Wooden Box,.

Description

Instrument Principle

Thermoplastic and Poly Vinyl Chloride substances when exposed to heat undergo many types of physical and chemical changes. The extent and type of change that takes place depends upon the severity of exposure of the sample to heat (i.e., temp. range), duration and rate of air flow. This test makes an assessment of change in Tensile Strength and Elongation of the material on subjecting them to accelerated ageing in hot air.

Standards

The Instrument is especially designed to suit the requirement of IEC: 60811-1-2, IS 10810 (Pt 10), IS 10810 (Pt 11)

Instrument

.

Each Cell of Thermal Ageing Oven is a complete instrument in itself i.e., fitted with individual Temp. Indicating Controller and an Hour Meter. This enables the user to test different type of samples at desired temperature and duration, depending upon test requirement of end users & testing authorities.

The Thermal Ageing Oven instrument consists of the following:

.

      A double walled chamber with inside chamber of thick aluminum sheet.

      Size of the chamber is 100 mm dia. x 300mm ht.

      Jacket type heater for the chamber.

      Digital Temp. Indicating Controller with sensor. Temp. range 0-200 ± 2 oC.

      Digital Flow Sensors with needle valves for controlling the volume of air passing through the chamber.

      Air inlet nozzle for the airline or air compressor.

      Specially designed split top cover to hold three to five test samples in each chamber

The Instrument can be operated with following Air configurations.

  1. Four Cell with 0-1 LPM air for 8-20 Air changes and other 4 cell with 0-10 LPM air for 100-200 Air Changes
  2. All Eight call with 0-1 LPM air for 8-20 Air changes
  3. All Eight Cell with 0-10 LPM air for 100-200 Air Changes..

To operate the Thermal Ageing Oven unit Compressed Air line is to be provided by the customer. If the customer is using his own air compressor, it may require one pressure regulator, pressure gauge & connecting pipes which are available from us.

Difference Between Hot Air Oven and Thermal Ageing Oven:

The test requirement is 8 to 20 air changes per hour, measured on a Flow Meter. According to IS standard, if the Oven does not have a regulated air supply, the oven is then termed as a Hot Air Oven and not a Ageing Oven (as offered by some manufactures) and cannot be used for ageing of samples.

After ageing in the ageing test oven for specified temp. & duration the samples are to be tested on Tensile Testing Machine for tensile & elongation. For preparation of Dumb-bell shaped samples, you may require hydraulic hot press, dumb-bell dies, dumb-bell cutting press, two-roll mill etc. Please ask for details.

 

IEC Panel Side Cells UL Panl Side
IEC Panel Side Cells UL Panel Side